| 
					
					
					 Log Calibrations Calibrations are an integral part of any log. If they were
					not run, or are now missing (or cut off the log), the
					analyst should use extreme care in picking log values.
					Standard calibrations for relatively modern logs (1960 to
					present) are documented in service company manuals, which
					should be part of your library. You will need dated versions
					of these manuals as the current manuals will show only
					current tools. This won't help for older logs in your files.
 Two
				calibration examples are shown below. The first is typical of
				analog log recordings (1960 - 1975+/-) and the second is typical
				of the digital log era. On analog calibrations, the calibration
				steps for several log curves are shown on  a log grid. For
				digital logs, a computer printout will show the calibration
				steps for various curves, as well as the permitted tolerance or
				range of allowable values. Calibrations before and after the
				survey, and in some cases shop calibrations are also attached. 
				Older logs (pre-1960) may have fewer, different, or no
				calibrations shown. Calibrations for different service companies
				vary slightly from each other, so be sure to obtain examples
				from all service companies which you normally use.  
				  
			  Since
			calibration details vary widely, it is impractical to publish all of
			them in a handbook of this type. It is strongly recommended,
			however, that you learn how to interpret and use calibration data,
			such as those shown in the illustrations in this section.
			Calibrations usually consist of low and high end points to define
			the log scale, and intermediate points to define linearity of scale.  
				Primary calibration of a log usually occurs under laboratory
				conditions or a test pit of known characteristics. Secondary
				calibration is a method for carrying primary calibrations to the
				service company field location by some device which simulates
				the laboratory readings. These are usually called shop
				calibrations. A third tier of calibration is a mechanism for
				transporting shop calibrations to the field for use at each well
				site. For
				example, a neutron log prototype is first calibrated in a test
				pit with known rock type and porosity. Then it is immediately
				run into a secondary calibrator of standard design, one of which
				will be available at each major logging center around the world.
				In the case of a neutron log, the secondary calibration is a
				tank of particular dimensions filled with diesel fuel. The
				readings in the secondary calibrator now constitute the main
				source of calibration. 
				Periodically thereafter, each tool is placed in the secondary
				calibrator, adjusted to read the correct response, and the field
				calibrator is placed on the tool. The tool response to this
				calibrator is then recorded. At each logging job, the tool is
				readjusted to read the same value when in the field calibrator
				environment. The field calibrator for a neutron log is a small
				gamma ray source at a short distance from the neutron log
				detector. This
				three stage process moves the primary calibration in the test
				pit in Houston to each well logged by the tool. Some logs
				require only a two stage calibration (such as induction logs)
				and some only require one stage (such as spontaneous potential
				or sonic travel time). 
				Calibrations performed before the log is run are called
				Before-Survey Calibrations, and those run after the job are
				called After-Survey Calibrations. Differences between Before and
				After calibrations need to be accounted for only if the
				difference is large enough to cause errors in the results of the
				log analysis. Even
				though the logging engineer tries to perform calibrations
				accurately and consistently, calibrations may be in error before
				the survey starts or may drift from their set values due to
				electronic problems. If these conditions prevail, the
				calibrations are said to be shifted. 
				  
				Several situations can arise if calibrations are clearly
				shifted. Both before and after survey, calibrations may be off
				by the same amount. Here, the log should be rescaled or a new
				scale constructed to correspond to the calibrations. Most
				computer aided log analysis software has a “block shift”
				function to do this. A
				drift may occur between before and after calibrations. Here the
				log must be rescaled at regular intervals to use up excess
				drift.
                 For
				example, assume a sonic log calibration showed the following: 
                
                  | 
                      
                        |  | Low End |  | High End |  
                        |  |  |  |  |  
                        | Before | 42 |  | 84 |  
                        | After | 43 |  | 82 |  
                        | Should have been | 40 |  | 80 |  |  If a
				40 to 140 scale was used for the logged interval, and the log
				was 3000 feet long, the following scales should be used: 
                
                  | 
                      
                        | Depth Interval |  | Scale |  
                        |  |  |  |  
                        | 3000 to 2000 feet |  | 42 to 147 usec/foot |  
                        | 2000 to 1000 feet |  | 42.5 to 145 usec/foot |  
                        | 1000 to 0 feet |  | 43 to 143 usec/foot |  |  
				These values are created by linear interpolation or
				extrapolation as required. Any log may be rescaled using linear
				algebra. A computer can apply a continuous, linear or non-linear
				shift as described by the user, providing the proper equation is
				incorporated. Most computer aided log analysis software has a
				“user-defined equation” function to do this type of
				re-calibration.  All
				induction resistivity and most laterologs logs should first be
				translated into conductivity, rescaled, and then translated back
				to resistivity. Most errors are in the sonde error setting,
				which is a linear shift in conductivity, not in resistivity.
				Calibrations may appear to be perfect, yet the log can read high
				or low in comparison to other logs in the area. Checkpoints for
				calibration shifts are the matrix base lines in clean,
				non-porous limestone, dolomite, or anhydrite, shale base lines,
				or overall position of the log curve with respect to another log
				in the same or nearby well in thick shale beds with good
				borehole conditions.
 |