| 
			
			
			 WATER ANALYSIS METHODS Laboratory water
			analysis is an essential measurement required for accurate water
			saturation calculations from log data. Water samples are collected
			from drill stem tests or produced fluids. In the case of produced
			fluids, the water should be captured from the flow line and
			separated later from the oil. Samples from separators or treaters
			may not be representative of formation water due to contamination.
			Samples from drill stem tests are usually taken at the top, middle,
			and bottom of the test recovery. The bottom sample should have the
			least contamination from drilling fluid invasion. The top sample
			will have the most contamination.
 
			
			Laboratories
			usually measure from 9 to 15 of the individual ions in a water
			sample, recorded in milligrams/litre (mg/l) or grams/cubic meter
			(g/m3). These two sets of units are equivalent: 1 mg/l =
			1 g/m3.  
			
			At low to moderate concentrations, one mg/l is very
			close to 1 part per million (ppm) so mg/l and ppm tend to be used
			interchangeably. The difference is that ppm = mg/l divided by the density of
			the water.
 
 In older reports, results were quoted in 
			
			grains per gallon (gpg)  One grain per US gallon equals 17.1 mg/l or
			approximately 17.1 ppm.
 
 The cations (positive ions) are measured by ion
			chromatograph. These are Sodium (Na), Potassium (K), Calcium (Ca),
			Magnesium (Mg), Iron (Fe), and sometimes Barium (Ba), Strontium (Sr),
			or Boron (Bo). The anions (negative ions) cannot be measured by ion
			chromatograph so Chloride (Cl) and sometimes Iodide (I) and Bromide
			(B) are still measured by titration. Bicarbonate (HCO3)
			and Carbonate (CO3) are calculated from the volume of
			acid required to reduce the pH to 8.3 (for CO3) and 4.5
			(for HCO3). Sulphate (SO4) is calculated by
			adding a volume of Barium Chloride and then measuring the turbidity
			of the solution. The sum of all of these measured ions becomes the
			calculated total dissolved solids (TDS).
   
			
			They also measure, in a routine analysis, the pH,
			relative density, and resistivity (Rw) of the water sample.  The Rw
			is measured in ohm-m and they will record the temperature at which
			it was measured. This temperature will be used to adjust the
			measured Rw to a common "laboratory temperature", usually 25 Celsius
			or 77 Farenheit. Some labs use different standard temperatures.   
			
			Some labs also provide mmol/l (or moles/m3). 
			mg/l (or g/m3) divided by the molar mass of the ion
			equals mmol/l (or moles/m3). Since mg/l is the same thing
			as g/m3, mmol/l is the same thing as mol/m3.
			Molar mass is derived from the atomic weight of the ion on the
			Periodic Table. For instance, the atomic weight of S is 32 and the
			atomic weight of O is 16, so the molar mass of SO4= 32 +
			4 * 16 = 96.   
			
			Most labs calculate milli-equivalents (MEQ).  MEQ
			equals mmol/l or moles/m3 multiplied by the valence or
			charge of the ion.  SO4 has a negative charge of 2 so 1
			mmol/l of SO4 equals 2 MEQ of SO4.  Sodium
			(Na) has a positive charge of 1, so 1 mmol/l of Na equals 1 MEQ of
			Na.     
			
			The logging company charts that convert ppm to Rw are
			based on ppm of pure NaCl. This is because the most common formation
			waters found in the world are NaCl based. These charts will give you
			the wrong Rw if the fluid is a mud filtrate which is mostly Na2SO4
			or if the fluid is Sodium Bicarbonate type formation water.   
			
			The diagram at the bottom of the water analysis is
			called a Stiff Diagram. It is a graphical representation of the
			different ions.  The shape of the Stiff Diagram can become a
			“fingerprint” which can allow us to distinguish whether the fluid is
			formation water or an introduced fluid, and can often distinguish
			the zone from which the formation water was produced.    
			
			
			 SAMPLE WATER ANALYSIS REPORT 
			 Water analysis report from a drill
			stem test recovery, showing chemical analysis, calculated and
			measured water resistivity, and Stiff diagram of chemical analysis.
			This is a fairly salty formation water with salinity of 146,000
			parts per million (ppm) and a resistivity of 0.066 @ 25C.
   
			
			
				
			 WATER SAMPLE CONTAMINATION Contamination of water samples by drilling fluid
			invasion is common and the Stiff Diagram helps to spot this
			problem. For instance, a typical gel-chem type mud filtrate recovery
			will have Na in MEQ divided by Cl in MEQ of 5 or greater. Most
			contamination problems require some experience and a supply of Stiff
			Diagram fingerprints that are reasonably consistent. 
			For more
			information on fingerprinting water recoveries, please contact Opus
			Petroleum Engineering Ltd.,
			
			www.opuspetroleum.com.
   
			The Canadian
			Water Resistivity Catalog is well screened, but field samples may be
			contaminated. The following rules of thumb are useful in detecting
			mud filtrate contamination or meteoric water recharge. 
			   1. Mg/l and
			ppm are approximately the same thing except in very saline waters. 
			   2. Most gel-chem
			mud filtrates are usually from 3000 to 8000 mg/l TDS. 
			   3. When the
			log header says gel-chem mud, they might mean gyp’ed-up mud. Gyp’ed-up
			mud usually has soda ash added to compensate for drilling through
			anhydrites. Gyp’ed-up mud filtrates are usually from 10,000 to
			25,000 mg/l TDS. 
			   4. KCl mud
			filtrates are usually from 30,000 to 50,000 mg/l TDS with lots of K
			and lots of Cl. 
			   5. Potassium
			Sulphate mud filtrates are usually from 50,000 to 80,000 mg/l TDS
			with lots of K and lots of SO4. 
			   6.
			Salt-saturated mud filtrates are usually 300,000 mg/l or higher. 
			   7. Generally
			speaking, formation waters increase in salinity with depth but see
			#12. 
			   8. Each zone
			should have a unique formation water “fingerprint” or Stiff Diagram
			unless it is hydraulically connected to another zone. 
			   9. This
			formation water “fingerprint” may change with location in the
			basin. 
			   10. Most
			formation waters have a milli-equivalent Na/Cl
			ratio of 0.6 to 1.2. 
			   11. Many
			formation waters are fresher than expected as they are affected by
			fresh water recharge from the surface. These recharge waters have a
			distinctive fingerprint which is high in bicarbonates and the milli-equivalent
			Na/Cl ratio is usually between 2 and 3. These fresh waters have been
			found in formations as deep as the Devonian and can under-run more
			saline formation waters. 
			   12. Never
			rely on one water analysis as being representative of the formation
			water in a particular pool and field. Try to find at least three
			water samples in your pool that have formation water
			characteristics, are close to the same TDS and have similar
			“fingerprints”. Then compare the Rw on your samples to the ones in
			the catalog. 
			  
			Agat Lab in Calgary uses a "Smart
			Chart", incorporating these rules,  to help identify clean
			water samples from those contaminated by mud filtrate or other
			chemicals. 
			
  This crossplot of TDS versus the Na/Cl ratio helps check for
			contamination. Data point 1 is in the formation water category.
			Point 2 is either naturally high in bicarbonates or contaminated by
			some gyp-based mud invasion. Point 3 is mostly gel-based mud
			filtrate.
 
 
			
			
					
				
			
			 Equivalent NaCl 
			 Water Salinity
			from Water Analysis The
				resistivity of a water sample can be calculated from its
				chemical analysis. To do this, an equivalent NaCl concentration
				must be determined based on the ionic activity of each ion.
 
			
			 Enter chart with total solids concentration of the sample in ppm
			(mg/kg) to find weighting factors for each ion present. The
			concentration of each ion is multiplied by its weighting factor, and
			the products for all ions are summed to obtain equivalent NaCl
			concentration.
 
 
			The math is
			pretty simple:1: TDS = SUM (IONi)
 2:
                WSe = SUM (IONi * FACTRi)
 
			Where:TDS = total dissolved solids (ppm)
 IONi = ion concentration of ith component (ppm)
 FACTRi = multiplier factor for ith component (ppm)
 WSe = equivalent NaCl concentration (ppm)
 
 
  NUMERICAL EXAMPLE: Assume formation-water sample analysis
 460 ppm Ca,
 1400 ppm SO4
 19,000 ppm Na plus Cl.
 
 Total dissolved solids concentration
 TDS = 460 + 1400 + 19,000 = 20,860 ppm
 Entering the chart with this total solids concentration
 Ca multiplier = 0.81
 SO4 multiplier = 0.45
 Na+CL multiplier = 1.00
 Equivalent NaCl concentration = 460 ´ 0.81 + 1400 ´ 0.45 + 19,000 ´
			1.0 = 20,000 ppm.
 
					
			
			 Water Salinity from
			Chloride Content Sometimes salinity is reported at the well site in ppm Chlorides
			instead of ppm NaCl equivalent.
 6: WSa = Ccl * 1.645
 Where:Ccl = water salinity (ppm Cl)
 WSa = water salinity (ppm NaCl)
 
				
				 COMMENTS: Use this relationship when chloride content of the water sample
				is known. Usually Cl content is derived at the
				well site from a drill stem test recovery. It is useful as a
				first approximation until the water sample is analyzed more
				accurately at a laboratory. The relationship is for pure NaCl
				solutions and the factor may be higher or lower if other ions
				are present.
 
				
				 NUMERICAL
                EXAMPLE: 1. Chloride concentration to salinity.
 WS = 11,600 ppm Cl * 1.645 = 19,000 ppm NaCl
 
					
			
			
			 Water Resistivity from Salinity
			AT ANY TEMPERATURE Crain's Model is used to convert a lab measured salinity
			to a formation water resistivity (RW) at any specific temperature (FT) in degrees Fahrenheit.
			The result is abbreviated as RW@FT
			throughout this Handbook. You can use equation 5 to convert a
			salinity to any arbitrary temperature, for example 75F or 77F
			(roughly 25C) to find the resistivity at laboratory conditions.
 1: FT = SUFT + (BHT - SUFT) / BHTDEP * DEPTH
 2: IF LOGUNITS$ = "METRIC"
 3: THEN FT1 = 9 / 5 * FT + 32
 4: OTHERWISE FT1 = FT
 5: RW@FT = (400000 / FT1 / WS) ^ 0.88
 
				
				 SALINITY FROM WATER RESISTIVITY Invert Crain's equation to solve for WS given RW at a a specific
				temperature FT1.
 6:
                WS = 400000 / FT1 / ((RW@ET) ^ 1.14)
 
 Where:
 BHT = bottom hole temperature (degrees Fahrenheit or Celsius)
 BHTDEP = depth at which BHT was measured (feet or meters)
 DEPTH = mid-point depth of reservoir (feet or meters)
 FT = formation temperature (degrees Fahrenheit or
				Celsius)
 FT1 = formation temperature (degrees Fahrenheit)
 RW@FT = water resistivity at formation temperatures (ohm-m)
 SUFT = surface temperature (degrees Fahrenheit or Celsius)
 WS = water salinity (ppm NcCl)
 
				
				 COMMENTS: Use this relation if salinity is known from laboratory measurements
				to obtain RW from lab data at any temperature.
 
 
  NUMERICAL
                EXAMPLE: 1. Salinity to water resistivity.
 RW@FT = (400000 / 102'F / 20,000 ppm) ^ 0.88 = 0.238 ohm-m @
                102'F
 (rounded to three significant digits)
 2.
				Water resistivity to salinity.WS = 400,000 / 102'F / ((0.250 ohm-m) ^ 1.14) = 19,000 ppm NaCl
 (rounded to three significant digits)
 
					
			
			 Water Resistivity from Salinity
			at Lab Temperature These models generate RW at laboratory temperature of 75F or 25C.
 Crain's Method (1974)
 1: RW@75F = (400000 / 75 / WS) ^ 0.88
 
 Bateman and Konen Method (1977)
 2: RW@75F = 0.0123 + (3647.5 / WS^0.955)
 
 Kennedy's Method (2015)
 3:  RW@75F = 1 / (24.30853 - 0.0364 *
			(0.1 * WS - 29.46515957) - 0.02922 * (0.1 * WS - 29.46515957)^2)
 
 
  SALINITY FROM WATER RESISTIVITY Crain's Method (1974)
 4:
                WS = 400000 / 75 / ((RW@75) ^ 1.14)
 
 Baker Atlas Method (2002)
 5: WS = 10 ^ ((3.562 - (Log (RW@75
			- 0.0123))) / 0.955)
 
 
  COMMENTS: For all
			practical purposes, the three models give the same RW value (see
			Graph 1 below). There are minor differences above 150,000 ppm NaCl
			which can only be seen when water resistivity is converted to water
			conductivity (see Graph 2 below).
 
 The effect on water saturation (SW%) is not very significant (+/-
			0.5% SW at low SW, +/- 2% SW at high SW, at 200,000 ppm NaCl).
 
 The 10 significant digits used in the Kennedy equation give a false
			sense of accuracy that is not warranted.
 
			 Graph 1: 
			Rw Models - Red line = Crain, Black line
			= Bateman and Konen, Blue line = Kennedy
 
 
  Graph 2: 
			Cw Models - Red
			line = Crain, Black line = Bateman and Konen, Blue line =
			Kennedy.
 The differences above 150,000 ppm NaCl have little impact on water
			saturation.
 
			
			 NUMERICAL
                EXAMPLE: 
  
 
					
			
			
			 RW from  Water Catalogs Water catalogues published by your local well logging society
                or similar catalogues created by searching in-house data bases
			are a necessary tool for well log analysis.
				A sample is shown below.
 
				 A sample of RW data from a water resistivity catalog, data is tabulated and also
 posted on
                a map, and is based on a standard temperature of 25 degrees Celsius
                (77 degrees Fahrenheit).
 
				 Water
				resistivity values in a catalog are recorded at a standard
				temperature, usually 75F or 77F (25C). Since water resistivity varies
				inversely with temperature, the catalog values must be
				transformed to a new value representing water resistivity at
				formation temperature (RW@FT) -- see
				next Section. 
 To use data from a water catalogue, it is usually necessary to
				do a little filtering. Nearly everything that can go wrong will
				raise the RW value recorded in the catalogue. Usually the
				minimum value from nearby offset wells is the best choice. It
				may be useful to gather all the values for the reservoir for a
				radius of 3 to 6 miles (5 to 10 Km) and prepare a histogram. On
				the histogram, find the point that represents the lower decile
				(10% of the data values are less than this value). Take the
				average of the data in this decile. You may want to eliminate
				obvious "impossible" values before you make the histogram.
 
 The
                following relationships are needed to manipulate water resistivity
                data prior to calculations of water saturation.
 1. Arps Method (1953)
 1:
                FT = SUFT + (BHT - SUFT) / BHTDEP * DEPTH
 2: KT1 = 6.8 for English units 
  KT1 = 21.5 for Metric units
 3: RW@FT = RW@TRW * (TRW + KT1) / (FT +
			KT1)
 4: RMF@FT = RMF@TRW * (TRW +
			KT1) / (FT + KT1)
 5: RMC@FT = RMC@STRW + (TRW * KT1) / (FT + KT1)
 
 Where:
 SUFT = surface temperature for temperature gradient
				(degrees Fahrenheit or Celsius)
 BHT = bottom hole temperature (degrees Fahrenheit or Celsius)
 BHTDEP = depth at which BHT was measured (feet or meters)
 DEPTH = mid-point depth of reservoir (feet or meters)
 FT = formation temperature (degrees Fahrenheit or Celsius)
 RMC@FT = mud cake resistivity at formation temperature (ohm-m)
 RMC@TRW = mud cake resistivity at surface temperature (ohm-m)
 RMF@FT = mud filtrate resistivity at formation temperature (ohm-m)
 RMF@TRW = mud filtrate resistivity at surface temperature (ohm-m)
 RW@FT = water resistivity at formation temperatures (ohm-m)
 RW@TRW = water resistivity at surface temperature (ohm-m)
 TRW = temperature at which water was measured (degrees Fahrenheit or Celsius)
 
				2. Hilchie Model (1984) ALL temperatures in Fahrenheit.6: KT1 = 10^ ( --0.340396 * log(RW@TRW)
			+ 0.641427)
 7:
                FT1 = SUFT + (BHT - SUFT) / BHTDEP * DEPTH
 8: RW@FT = RW@TRW * (TRW + KT1) / (FT1 +
			KT1)
 9: RMF@FT = RMF@TRW * (TRW +
			KT1) / (FT1 + KT1)
 10: RMC@FT = RMC@STRW + (TRW * KT1) / (FT1 + KT1)
 
 Where:
 FT1 = formation temperature (degrees Fahrenheit ONLY)
 
 COMMENTS:
 Use
			these relations when RW@TRW
				is known from measured data. This transformation can be made on
				the chart below. The Hilchie model accounts for the slight
			curvature at low and high temperatures on the chart, but Arps model
			is quite sufficient for practical petrophysics.
 
			
			 NUMERICAL
                EXAMPLE: 1. Water resistivity at formation temperature.
 English units example:
 RW@FT = (0.32 ohm-m @ 77'F) * (77 + 6.8) / (102 + 6.8) = 0.25
                ohm @ 102'F
  Metric
                units example:RW@FT = (0.32 ohm-m @ 25'C) * (25 + 21.5) / (39 + 21.5) = 0.25
                ohm-m @ 39'C
 
				
				 Schlumberger Chart GEN-9: Water resistivity - Temperature - Salinity relationships
 
 
  ESTIMATING
				SURFACE AND FORMATION TEMPERATURES Temperature
				measurements specific to your area of interest are going to
				sparse and you will have to do some searching for useful data.
				The map below gives a good idea of what to use for surface
				temperature (SUFT). Temperature versus depth data from log
				headings can be plotted to estimate a best fit temperature
				gradient line. It doesn't have to be a straight line. See
				representative graphs below. In areas with sparse data, use the
				temperature gradient maps supplied below.
 
				
			 Mean Annual Surface Temperature Map for North America - degrees F
 The 40 deg F contour follows roughly along the US-Canada border
			except along the coast lines.
 Contour interval is 5 deg F.
 
				 Typical Depth - Temperature profiles. Create your own by
				best fit of BHT vs BHTDEP from data on log
 headings or DST reports.
 
			 Temperature gradient for USA -
			degrees Celsius per 1000 meters,
 North American Heat Floe Map
			(3 MB)    
			Legible Legend for NA Map
 
			 Temperature at 5000 meters for
			Australia
 
					
  META/LOG "RW" SPREADSHEET -- Water Resistivity Calculations This spreadsheet calculates RW at formation temperature using 5
				different methods.
 
 Download this spreadsheet:
 SPR-07 META/LOG WATER RESISTIVITY (RW) CALCULATOR
 Calculate water resistivity (RW),
						5 methods,
 
 
			 Sample of "META/RW" for calculating water resistivity from various
			methods.
 
 |